TestProcess.Measure.Ratio.FrequencyModulation.Delta.Rate

A test process measuring the deviation change of a frequency modulated signal as the modulation rate frequency is changed from the referenced modulation rate frequency.  The signal source is set up at a specified fixed carrier frequency, with a fixed deviation frequency then only the modulation rate frequency is changed from the reference rate frequency to the test rate frequency.  The resulting values are measures as a dB change or % change.

Required Parameters

  • CarrierFrequency – RF Carrier Frequency
  • Deviation – Frequency Modulation Deviation
  • RefRate –  Reference Modulation rate Frequency, usually 1 kHz
  • Rate –  Modulation Rate Frequency

Optional Parameters

  • Power
  • HiPassFilter
  • LowPassFilter
  • Detector
  • UUT.Output – Output Name of the UUT Connecting Point

Output Value & Uncertainty

  • dB – Measured Value – Reference Value
  • % – Reference Value / Measured Value * 100